Actimetrics

Phone:
765-423-1505

LumiCycle 32 Specifications

Tissue holders 35 mm dishes (Falcon)
Capacity 32 dishes
Detectors 4 Photomultiplier tubes (photon counting mode)
Each tube records alternately from 8 dishes
Dark counts <10/sec @ 20 deg. C
50/sec typical @ 35 deg. C
Optical Filters
for color separation
Diameter: 25 mm
Thickness: 3 mm
Long Pass (from Edmund Optical)
Short Pass (from Edmund Optical)
Operating temperature Up to 37 deg. C – ambient humidity
(as set by incubator or environmental room)
Dimensions 14″ W x 13.5″ H x 14″ D (35.5 x 34.3 x 35.5 cm)
18 kg
Front loading
Count period 10 s – 1 hr program selectable
System requirements Windows 7/8/10
1 USB port
1280 x 1024 resolution monitor recommended
Analyses Actogram
Periodogram
Spectral (sinusoidal fit and damped sinusoidal fit)
Phase shift (double sinusoidal fit to defined periods)
Baseline subtraction (low-order polynomial)
Warranty One year, parts and labor